English | Español | Others:
Home   Products   Solutions   Services   Inquiry   About Us   Contact Us  
Medical Equipment
Laboratory Equipment
Homecare Product
  Veterinary Equipment
 
  Product Details
Product
Anaesthsia System
Code ID: AM-301D
Model ID: AM-301D
Certificate: ISO, CE
Inquire
Print
 
Feature:

Anaesthsia System(AM-301D)
Model: AM-301D
 
Main feature:
◆ Be used for closed、semi-closed、semi-open inhalation anaesthesia, controlled or assisted ventilation, monitoring and display of the ventilation parameters.
◆ Applied for adult and child.
◆ Compacted breathing system meets the standard of ISO 8835-2. Good airtight, easy to operation, clearance and sterilization.
◆ Vaporizer meets the standard of ISO 8835-4. Accurate concentration, temperature、flow and pressure compensation technology. Various options of Isoflurane, Enflurane, and Sevoflurane.
◆ Anaesthesia ventilator meets the standard of ISO 8835-5.
Gas driven and electrically controlled
Big TFT screen
IPPV, SIMV, PCV ventilation mode
Inhalation platform and PEEP function
Volume and pressure control
Compliance compensation function
Alarm, indication and protection functions for high pressure、leadoff and apnea
Built-in battery for backup power source
Main technical specification:
●Power : AC 22OV, 50Hz
●Gas: O2, N2O, Air      Working pressure: 0.28MPa~0.6MPa
● Flowmeter
――Adjusting range
O2 flowmeter (Double tubes): 0.1~10L/min
N 2O flowmeter (Double tubes): 0.1~10L/ min      
Air flowmeter (Single tube): 0.1~10L/min
――Interlock device to ensure the O2 concentration not less than 25%.
● Vaporizer: (Isoflurane, Enflurane, sevoflurane.  Double or single vaporizer option)
Range of concentration: 0~5%
Temperature、flow and pressure compensation technology
●Adjusting range of main ventilation parameters
Frequency: 1/min~100/min
I/E Ratio: 4:1~1:8
Tidal Volume: 0L~1.5L
●Ventilation parameter monitoring:
Digital display: tidal volume, minute ventilation, frequency, airway pressure, O2 Concentration, lung compliance, and waveforms.
● Alarm, indication and protection for;
Power off, low voltage of battery, minute ventilation beyond the limit, pressure beyond the limit, O2 concentration beyond the limit, Apean, and insufficiency of O2, etc.

 
Configuration:
 
Home   Products   Solutions   Services   Inquiry   About Us   Contact Us  

Shinova Systems Co., Ltd. / Shinova Medical Co., Ltd. ©2000-2016