Descriptions
Suitable for experimental animals such as rats, mice, rabbits, cats, hamsters, rats guinea pigs, etc.Coaxial tube gas circuit design saves operating space and is easy to operate The internal serpentine pipeline design for clinical application ensures that the gas output is not affected by pressure changesOxygen or air can be selected as gas supply sourceProvide accurate oxygen flow meter; flow controllable range is 0-500ml/min The anesthesia vaporizer has stable output, good airtightness, safety and reliability, and is not affected by changes in flow, temperature, flow rate, and pressureThe vaporizer has automatic flow and temperature compensation, the accuracy is not lower than 0.5%, and the temperature is maintained in the range of 15-35°C, which reduces the number of dosing and makes it more convenient to useAnti-accidental opening lock structure and closed state safety protection structure to ensure safetyAccording to experimental requirements, different types and sizes of animals can choose various specifications and accessories(such as induction box,anesthesia mask ,etc)With exhaust gas recovery system to ensure a clean laboratory environmentOne-key switch of gas path, saving time and effortWith fast oxygen function to ensure that animals inhale oxygen Can be upgraded to mobile, wall-mounted anesthesia machine.
Specifications
Model
AneLab-10
Gas source supply flow rate
0.2~15L/min,If the concentration is greater than 4%,0.2~8L/min
Gas source supply quality requirements
Medical pure oxygen and air mixture, or O2 and N2O mixture oxygen and air
Channel
1
Control concentration range
Sevoflurane : 0.2-8% or Isoflurane : 0.2~5%(v/v)
Control knob control value
Sevoflurane: 0-0.2-0.5-1.0-1.5-2.0-2.5-3.0-3.5-4.0-4.5 -5.0-5.5-6-6.5-7-7.5-8 %(v/v)
Isoflurane:0 - 0.2 - 0.5 - 1.0 - 1.5 - 2.0 - 2.5 - 3.0 - 3.5 - 4.0 4.5-5.0%(v/v)
Concentration control accuracy
±5% of full scale or ±15% of set value
Anesthetic dose
The total volume of the dry wick is about 300mL,The visible liquid level is about 250mL between the minimum and maximum scale.
Consumption of anesthetic
About 3 x supply gas flow (L/min) x set concentration value (%(v/v)) ,When the isoflurane concentration is 2% and the gas flow rate is adjusted to 600mL/min, a 100mL bottle of isoflurane can be used for approximately 28-30 hours.
Loss of anesthetic
22℃, 0 graduation, less than 0.5mL/24h.
Maximum pressure load
20kPa
Maximum use tilt angle
30°
Material
Mainly aluminum
Working conditions
temperature :5~40℃,Atmospheric pressure :700~1060hPa
Storage conditions
temperature :-40~55℃,Atmospheric pressure :500~1060hPa